ตู้ควบคุมอุณหภูมิและความชื้น Constant climate chamber with humidity

Constant climate chamber with humidity

The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate conditions. From programming to documentation, this constant climate chamber meets all critical requirements.

  • Electronically controlled APT.line™ preheating chamber and refrigerating system assuring temperature accuracy and reproducible results
  • Temperature range 0 °C to 70 °C (without humidity)
  • Temperature range 10 °C to 70 °C (with humidity)
  • Humidity range 10 % to 80 % RH
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
    • User friendly LCD screen
    • Easy-to-read menu guide
    • Integrated electronic chart recorder
    • Variety of options for the graphic display of process parameters
    • Real time clock
  • Electronically controlled humidification and dehumidification system with capacitive humidity sensor
  • Suitable for stability tests according to ICH guideline Q1A
  • Inner glass door with sealing
  • Independent adjustable temperature safety device class 3.1 (DIN 12880), with visual and audible temperature alarm
  • Access port with silicone plug Ø 30 mm, left side
  • Safety connection kit for water supply and drainage, including water hose up to 1 m Height
  • Ethernet interface for GLP/GMP and the FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControl-System software
  • 2 stainless steel racks
  • BINDER test confirmation
KBF 115 (E5.2)
Exterior dimensions
Width (mm) 885
Height (incl. feet) (mm) 1050
Depth (incl. door handle, I-triangle, connection 80 mm) (mm) 730
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 100
Steam space volume (l) 156
Number of doors (ea.) 1
Inner glass door(s) (ea.) 1
Interior dimensions
Width (mm) 600
Height (mm) 483
Depth (mm) 351
Interior volume (l) 102
Racks (number standard/max.) 2 / 5
Load per rack (kg) 30
Permitted total load (kg) 100
Weight (empty) (kg) 129
     
Temperature data (without humidity)
Temperature range (°C) 0 - 70
Temperature variation at 25 °C (± K) 0,2
Temperature variation at 40 °C (± K) 0,2
Temperature fluctuation (± K) 0,1
Max. heat compensation up to 40 °C (W) 200
     
Climate data (with humidity)
Temperature range (°C) 10 - 70
Temperature variation
at 25 °C / 60 % RH (± K) 0,2
at 40 °C / 75 % RH (± K) 0,2
Temperature fluctuation
at 25 °C / 60 % RH (± K) 0,1
at 40 °C / 75 % RH (± K) 0,1
Humidity range (% RH) 10 - 80
Humidity fluctuation
at 25 °C / 60 % RH (± % RH) ≤ 2
at 40 °C / 75 % RH (± % RH) ≤ 2
Recovery time after door was opened for 30 sec
at 25 °C / 60 % RH (min.) 15
at 40 °C / 75 % RH (min.) 8
     
Electrical data
Voltage (± 10 %) 50 / 60 Hz (V) 200 - 240, 1 N ~
Nominal power (kW) 2,0
Energy consumption 40 °C / 75 % RH (W) 1) 470
Noise level approx. (dB (A)) 52

Temperature-humidity chart

A: Guaranteed condensation free range
B: Temporary usable operation (up to 24 h)

Heat compensation