Constant climate chamber with additional photometry

Constant climate chamber with additional photometry

The special advantage of the BINDER KBF LQC series: the patented photometry for photostability testing.

Two spherical 3D ball-type sensors capture visible light intensity and UV light intensity – if the desired light dosage is achieved,

the KBF LQC constant climate chamber switches off automatically.

KBF LQC 240 (E5.2)
Exterior dimensions
Width (mm) 930
Height (incl. casters) (mm) 1460
Depth (incl. door handle, I-panel, connection 80 mm) (mm) 880
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 100
Steam space volume (l) 348
Number of doors (ea.) 1
Inner glass door (ea.) 1
     
Interior dimensions
Width (mm) 650
Height (mm) 785
Depth (mm) 485
Interior volume (l) 247
Racks (number standard/max.) 2 / 7
Load per rack (kg) 30
Permitted total load (kg) 100
Weight (empty) (kg) 213
     
Temperature data (without humidity)
without illumination cassettes (°C) 0 - 70
with illumination cassettes, with illumination (°C) 10 - 60
Max. heat compensation up to 40 °C with illumination (W) 400
     
Climate data (with humidity)
Temperature range without illumination cassettes (°C) 10 - 70
Temperature range with illumination cassettes with illumination (°C) 10 - 60
Temperature variation with illumination
at 25 °C and 60 % RH (± K) 0,6
at 40 °C and 75 % RH (± K) 0,6
Temperature fluctuation with illumination
at 25 °C and 60 % RH (± K) 0,2
at 40 °C and 75 % RH (± K) 0,2
Humidity range without illumination cassettes (% RH) 10 - 80
Humidity range with illumination cassettes, with illumination (% RH) 10 - 75
Humidity fluctuation with illumination
at 25 °C and 60 % RH (± % RH) 1,5
at 40 °C and 75 % RH (± % RH) 2
Recovery time after doors were open for 30 sec.
at 25 °C and 60 % RH with illumination (min.) 3
at 40 °C and 75 % RH with illumination (min.) 4
Illumination data per light cassette
ICH-compliant illumination device for photostability testing (Lux) / (UVA W/m²) 7500 / 1,4
     
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage (± 10 %) 50 Hz (V) 230, 1 N ~
Nominal power (kW) 2,4
Energy consumption at 40 °C / 75 % RH 1) (W) 660
Noise level approx. (dB (A)) 52

BINDER Light Quantum Control (LQC)

2 spherical sensors for non-directional measurement, directly at the sample location

Temperature-humidity chart without illumination

A: Guaranteed condensation free range / B: Temporary usable operation (up to 24 h)

Temperature-humidity chart with illumination

A: Guaranteed condensation free range / B: Temporary usable operation (up to 24 h) / C: Deviation of technical data may be possible

Heat compensation without illumination

Heat compensation with illumination

Light spectrum