MP513 Lab Conductivity Meter แบบตั้งโต๊ะ

MP513 Lab Conductivity Meter

For normal solution for measuring the conductivity, especially in the areas of education and research used.

 

INTRODUCTION

 

Meter meets with the requirement of international GLP standards. Its features up to automatic calibration, automatic temperature compensation, data storage, timing measurement, clock display, RS-232 output, function setting and self-diagnose etc.

There are three different measurement modes: stable display mode; time measurement mode; continuous measurement mode.

Equips Conductivity/ATC electrode. Its measurement result is fast and precise.

Adopt advanced and original conductivity measurement technology, use 1 pc of conductivity electrode (K=1), it can meet measuring accuracy requirement for range within 0.5μS/cm ~ 200mS/cm by using only one-point calibration.

Ability to switch among conductivity, TDS, salinity and resistivity. Multinomial calculation for TDS and salinity to ensure the conversion precision of the full scale.

Meter has the only product serial number.

Dustproof and waterproof meter meets IP54 standards. All sockets are seal protected by the silica gel caps.

 

SPECIFICATIONS

 

Conductivity

Measuring range

 

 

 

 

 

Resolution

Accuracy

Temp. compensation range

Electrode constant

Reference temperature

Conductivity:(0.00~20.00μS/cm 20.0~200.0μS/cm

200~2000μS/cm  2.00~20.00mS/cm

20.0~200.0mS/cm

Resistivity:(0 ~ 100MΩ·cm

TDS:  0 ~ 100g/L

Salinity 0 ~ 100ppt

0.01/0.1/1μS/cm  0.01/0.1/1 mS/cm

Meter:±0.5% FS,connect meter with electrode:±0.80% FS

0 ~ 50)℃(automatic

0.1 / 0.5 / 1 / 5 / 10 / 50 / 100  cm-1

25℃、20 and 18

Temperature

Measuring range

Resolution

Accuracy

-10~ 110

0.1

5~ 60range±0.4  other range±0.8

Other Technical Parameters:

Data storage

Storage content

 

Power

Communication connector

Size and weight

Quality and safety certification

300 groups

Measuring value series number, measuring value, temperature, ATC or MTC state, measuring date and time

DC9V/300mA

RS232

160 × 190 × 70mm/880g

ISO9001:2008, CE and CMC

Working Condition

Environment temperature

Environmental humidity

IP rating

5 ~ 35 ℃(0.01grade 15 ~ 30 ℃(0.001grade

≤75%

IP54 Dustproof and waterproof

 

STANDARD KIT INCLUDES

 

1. Model MP513 lab conductivity meter

1 unit

2. Model 602 flexible electrode holder

1pc

3. 2301T-M plastic pH/ATC three-in-one combination electrode

1 pc

4. 1413μS/cm conductivity standard solution (50mL)

2 bottles

5. 9V power adapter

1pc

6. RS-232 communication cable

1pc

7. MP513 communication software CD

1pc

8. Operation manual

1 unit

9. Brief operation manual

1 unit

10. Packing List

1 pc